Oxygen Concentrator SIM Travelair Special

 
This complex combination of technical characteristics was considered indispensable to create a device that would be complementary to home oxygen therapy systems such as the cryogenic containers (base) and above all, its respective "stroller", emphasising the advantages already known. In fact, while the "stroller" allows a series of exclusive advantages for the patient, it was developed to offer autonomy limited to a few hours, and therefore not adapt to permit lengthy travel by car, vacations and/or occasional overnight stays away from home.

The TRAVELAIR® is normally used for lengthy trips by car, in camper, by boat, overnight stays at hotels, at summer homes, on cruises, etc., thus realising that technological dream of more than ten years ago created by a prevalently Italian group of obstinate doctors and engineers.

Small: only 44,5 cm. tall

Portable: weighs only 12 Kgs.

Reliable: 2-years or 1000 working hour guarantee

Offers 2 continuous flow settings: 1 and 2 lpm respectively

Offers 6 demand flow settings: 2.5 - 3 - 3.5 - 4 - 4.5 and 5 lpm respectively

Simple to use: completely automatic

Maximum safety: oxygen monitor, cannula alarm, O2 flow indicator, high temperature indicator, dust filter blocked indicator

The only Oxygen Concentrator constructed to work on 13V d.c.± 1 V d.c. power and on 110-230Va.c. through its internal switching power supply.

Includes a special cable for use in the car/RV/boat via the cigarette lighter, rechargeable battery (1 hour autonomy approx.) and battery charger available.

Technical specifications REF. TRAVELSPEC

O2 concentration ± 3%*:
1 • 2 LPM - >90% "CONTINUOUS" setting

2,5 • 3 • 3,5 • 4 • 4,5 • 5 LPM - >90% "DEMAND" setting

Clinical data:
Breaths per minute (BPM) ml/pulse

25 Min 50 – Max 75

Maximum system pressure:
25 psi ( 1,7 bar)

Oxygen outlet pressure:
6 psi (0.4 bar) "CONTINUOUS" setting

20 psi (1.4 bar) "DEMAND" setting

Electrical data:
13 Volts c.c. ± 1 Volt c.c. 110-230V a.c. - 50/60Hz

Fuse:
1 fuse mod. 5 x 20, 4 Amp.

Requirements for use in the car:
The cable from the cigarette lighter to the car battery must be 4 - 6 mms. in diameter.

Maximum current:
14 Amp. - 180 Watts - 13 Volts c.c.

1 Amp. - 230 Watts - 230 V a.c.

Operating temperature:
+10°C - +40°C

Dimensions (HxWxD):
445 x 290 x 265 mm. - 17.5 x 11.4 x 10.4 inch

Dimensions with handle completely extended (HxWxD):
975 x 290 x 265 mm. - 38.4 x 11.4 x 10.4 inch

Weight:
12 Kg. - 26,4 lbs.

Noise:
58db

Indicators:
ON-OFF switch, cannula alarm, O2 flow, oxygen monitor, high temperature, blocked filter, hourmeter, continuous/demand setting, external battery discharged.

Audible alarms:
Electrical power insufficient, low oxygen concentration <82%, the unit becomes disconnected from any power source,

Filters:
Dust filter (1), compressor inlet filter HEPA (1), product filter (1).

Movement and transport:
Two robust rear wheels. Telescoping handle.

Guarantee:
2 years or 1000 Hours (whichever comes first)

Safety standards:
EN 60601-1 (1998), EN 60601-1-2 (2001)

Device classification:
MDD 93/42/CEE: CLASS IIa

CEI 62-5 (1991): CLASS I type BF

* these concentrations were obtained under optimum test conditions between 0 and 1000 mt. above sea level - Rel. Hum. 15-95%

 